Abstract

Gentamicin levels were determined in 100 serum specimens by a new latex agglutination inhibition card test, a radioimmunoassay (RIA), and a bioassay. Correlation coefficients determined by linear regression analysis demonstrated that the levels obtained by the latex agglutination inhibition card test had a high degree of correlation with the RIA and could be performed much faster and more economically when processing small numbers of specimens. The bioassay had a slightly lower degree of correlation with both the RIA and the latex test and was adversely influenced by concurrently administered antibiotics which could not be eliminated by beta-lactamase. When measuring gentamicin concentrations above 2 micrograms/ml, the coefficient of variation was less than 14% for the latex agglutination assay compared with 15% for the bioassay and 12% for RIA. The latex agglutination inhibition card test is a rapid, accurate, specific, and reproducible method for monitoring gentamicin levels in patients and is particularly applicable for laboratories processing small numbers of specimens.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call