Abstract

We report the results of a bicuspid expanded polytetrafluoroethylene (ePTFE) valved conduit used for right ventricular outflow tract reconstruction (RVOTR). Between November 2005 and February 2009, 12 conduits were used for RVOTR. The mean age and weight of patients were 43.5 ± 46.4 months and 13.4 ± 8.6 kg. The main diagnosis was tetralogy of Fallot with pulmonary atresia in eight patients. The most common conduit size was 18 mm. The mean follow-up was 88.0 ± 35.9 months. There were no operative and late mortalities. At discharge, the mean peak systolic pressure gradient across the RVOT was 14.1 ± 11.3 mmHg. There was no conduit valve regurgitation in nine patients. At the latest echocardiography (mean follow-up: 84.3 ± 35.5 months), the mean peak systolic pressure gradient across the RVOT was 59.7 ± 20.2 mmHg, and there was no conduit valve regurgitation in six patients. Freedom from conduit malfunction was 100% and 83.3%, at 1 and 8 years, respectively. Two conduits were explanted due to sternal compression and four from conduit malfunction. Freedom from explantation was 83.3% and 74.2% at 2 and 8 years, respectively. ePTFE bicuspid valved conduit has good late function in terms of valve regurgitation, but the pressure gradient across the conduit increases with time, which is the main cause of conduit failure and explantation.