Abstract
We present a wavelength meter with picometer-scale resolution based on etaloning effects of inexpensive glass slides and the built-in color filters of a consumer grade CMOS camera. After calibrating the device to a commercial meter, we tested the device's calibration stability using two tunable visible lasers for a period of over 16 days. The wavelength error over that entire period has a standard deviation of 5.29 parts per million (ppm) about a most probable error of 0.90 ppm. Within 24 hours of calibration, this improves to 0.04 ppm with a standard deviation of 3.94 ppm.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.