Bioaerosol mass spectrometry is being developed to analyze and identify biological aerosols in real time. Characteristic mass spectra from individual bacterial endospores of Bacillus subtilis var. niger were obtained in a bipolar aerosol time-of-flight mass spectrometer using a pulsed 266-nm laser for molecular desorption and ionization. Spectra from single spores collected at an average fluence of approximately 0.1 J/cm2 frequently contain prominent peaks attributed to arginine, dipicolinic acid, and glutamic acid, but the shot-to-shot (spore-to-spore) variability in the data may make it difficult to consistently distinguish closely related Bacillus species with an automated routine. Fortunately, a study of the laser power dependence of the mass spectra reveals clear trends and a finite number of "spectral types" that span most of the variability. This, we will show, indicates that a significant fraction of the variability must be attributed to fluence variations in the profile of the laser beam.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.