Abstract

Capillary electrophoresis coupled with sheath-flow laser-induced fluorescence (LIF) detection has been shown to offer outstanding sensitivity for chemical and biochemical analysis. However, a major drawback remains with the complexity of the optical configuration traditionally employed. Here we present a simplified confocal optics based on fiber optics and micro gradient-index (GRIN) lenses for modular optical design in capillary electrophoresis with laser-induced fluorescence. We demonstrate the use of the optical system with a sheath-flow cuvette as the laser-induced fluorescence detector for capillary electrophoresis. The system's performance was established with concentration detection limits of 8±2pM and mass detection limits of 57 zeptomole for a standard sodium fluorescein sample.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.