Abstract

The paper presents the application of laser-induced breakdown spectroscopy (LIBS) for the on-line multielement analyses of glass melts in a vitrification process of high level liquid waste (HLLW). The third harmonic pulse of an Nd:YAG laser is used for the generation of plasma on the molten glass surface and the plasma emission is monitored by an echelle spectrometer with an intensified charge-coupled device (ICCD), which simultaneously covers the wavelength range from 200 to 780 nm. Twelve different reference HLLW glass melts with a complex composition of about 27 chemical elements are simulated on a laboratory scale, varying the HLLW component concentration. By real-time analyses of the reference glasses at 1200 °C, the analytical method is calibrated. A multivariate regression approach with partial least squares (PLS) is used for the data evaluation. The LIBS method thus calibrated is then applied for the multielement analysis of molten glass samples from the prototype vitrification plant under operation in our institute. The results underline that the LIBS method can be applied to a vitrification process for the on-line multielement analysis of highly radioactive glass melts.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.