Abstract

Time-resolved laser-induced breakdown spectroscopy was applied for quantitative measurement of lead content in concrete at levels down to 10 ppm. The breakdown was formed at the sample surface by a Q-switched ND:YAG laser operating at a 1.06-μm wavelength and a repetition rate of 10 Hz. Contamination levels were inferred from the ratio of the integrated emission line of lead to a known reference line of the matrix. The lead contamination can be determined on an absolute scale down to 10 ppm at an optimum delay time of 3.0 μs. These results were derived from analysis of the temporal evolution of the calibration function within a 0.1- to 19.0-μs time range. The calibration function exhibits no dependence on the incident laser pulse energy, which was varied from 250 to 400 mJ.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.