Abstract

Objective, non-invasive examination, techniques in addition to clinical parameters, are required to follow-up the wound healing of flaps. With the new laser Doppler Scanner (LDI DIM 1.0 Lisca Development AB, Sweden) it is possible, for the first time, to measure and image the microcirculation continuously, non-invasively and without contact with the wound, in an area of 12 cm square maximum. We performed measurements and simultaneous two-dimensional imaging of the microcirculation 24, 48, 72 h and 5 and 14 days postoperatively in 20 patients, who had had reconstruction procedures performed using random or axial pattern flaps. The perfusion diagrams were correlated to the clinical appearance. Necrotic areas, venous stasis and normal course of wound healing can be clearly visualized and differentiated from one another. The new laser Doppler imaging system seems to be an excellent aid for following up and planning of flaps in plastic and reconstructive surgery.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.