Abstract

The paper aims to compare the results obtained with the same laser source with a large beam and with small beams. These latest were shaped from phase objects implemented to obtain several small beams from a single larger beam. The consistency of the results from both sets of measurements is shown. It validates the assumptions made and the specific mathematical treatments implemented to establish the link between the two approaches. It also validates and strengthens the approach developed from the rasterscan procedure used to measure damage densities from the scanning with beams of small dimensions. This shows that small beam tests are reasonably representative of tests carried out with large beams.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.