Abstract

In recent years there has been a growing awareness of the possible hazards caused by anaesthetic gases in operating theatres. The laryngeal mask airway provides an alternative both to tracheal intubation and the face mask although the implications for operating theatre contamination have not been quantified. This paper describes the incidence and magnitude of exposure of theatre personnel to waste anaesthetic gases during laryngeal mask airway anaesthesia. The leakage of anaesthetic gases to the anaesthetist's breathing zone was monitored using a Bruel & Kjaer Multi Gas Monitor, Type 1302 during 50 general anaesthetics employing either spontaneous (n = 24) or controlled (n = 26) ventilation. All patients were anaesthetised with propofol, alfentanil and nitrous oxide. There was no statistically significant association between the amount of anaesthetic gas leakage and ventilation method. The laryngeal mask airway meets occupational safety requirements on nitrous oxide concentrations in the operating theatre environment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.