Abstract

We present a large-area perfect blackbody sheet, which would offer a planar standard radiator for high-precision thermal imager calibration. Polydimethylsiloxane (PDMS) sheets with nano-precision surface micro-cavity structures achieve both ultra-low reflectance (ultra-high emissivity close to unity) over the thermal infrared wavelengths and high durability to mechanical contact. The investigation on the geometrical parameters of the conical micro-cavities, that is, radii and aspect ratios (ratio of height to radius), confirmed that the PDMS blackbody sheet with a micro-cavity radius of ∼6 µm and an aspect ratio of ∼4 exhibits the optimum hemispherical reflectance of less than 0.002 (emissivity of higher than 0.998) at the thermal infrared wavelengths (6-15 µm). Furthermore, the large-area PDMS blackbody sheet of 100 mm × 80 mm maintained an excellent in-plane uniformity of the emissivity. This unprecedented large-area perfect blackbody conforms to the International Electrotechnical Commission (IEC) standard recommendation regarding thermal imager calibration for fever screening in terms of the emissivity performance.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.