Abstract
Analyze energy-induced damage to the swine vagina during laparoscopic hysterectomy. Laparoscopic colpotomy was performed in swine using ultrasonic, monopolar, and bipolar energy. Specimens (n = 22) from 13 swine were stained with hematoxylin and eosin and Masson's trichrome for energy-related damage. The distal scalpel-cut margin was used as reference. Energy induced damage was assessed by gynecologic and veterinary pathologists blinded to energy source. Injury was most apparent on Masson's trichrome, demonstrating clear injury demarcation, allowing consistent, quantitative damage measurements. Mean injury was 0 ± 0 μM (scalpel, n = 22), 782 ± 359 μM (ultrasonic, n = 7), 2016 ± 1423 μM (monopolar, n = 8), and 3011 ± 1239 μM (bipolar, n = 7). Using scalpel as the reference, all were significant (P < .001). All energy sources demonstrated tissue damage, with ultrasonic showing the least and bipolar the greatest. Further study of tissue damage relative to cuff closure at laparoscopic hysterectomy is warranted.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.