Abstract

Electrical stimulation (ES) is a promising therapy for multisegmental gastrointestinal (GI) motility disorders such as gastroparesis with slow-transit constipation or chronic intestinal pseudo-obstruction. Wireless communicating GI devices for smart sensing and ES-based motility modulation will soon be available. Before placement, a potential benefit for each GI segment must be intraoperatively assessed. A minimally invasive multisegmental electromyography (EMG) analysis with ES of the GI tract is required. Two porcine experiments were performed with a laparoscopic setup. Multiple hook-needle electrodes were subserosally applied in the stomach, duodenum, jejunum, ileum, and colon. EMG signals were acquired for computer-assisted motility analysis. Gastric ES, duodenal ES, jejunal ES, ileal ES, and colonic ES were applied. Further technological and rapid regulatory solutions are desired to initialize a clinical trial of the next generation devices in the near future. We demonstrate a laparoscopic strategy with EMG analysis and ES of multiple GI segments. Thus, GI function may be evaluated before theranostic devices are placed. Extended GI resection or organ transplantation may be delayed or even avoided in affected patients.