Abstract

In this paper we describe the optimization of transmission X-ray targets by Monte-Carlo simulation for a laboratory X-ray microscopy setup. We identified two optimal target layer thicknesses (0.1 μm and 0.7 μm) for a high-resolution target and a high-flux target. Measurements show a decrease in focal spot size by one third or an increase in X-ray flux by a factor of three compared to those of a standard micro-focus target. Focal spot sizes down to 154 nm and 260 nm are achievable with the optimized targets. Simulation results for the X-ray flux match well to the experimental results, whereas the results for the focal spot sizes still show discrepancies attributed to the simplified simulation setup.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.