Abstract

Gas-to-particle conversion takes readily place in the atmosphere. Detecting the initial clusters, which act as embryos for the newly formed particles, is beyond traditional aerosol instrumentation. Charged atmospheric clusters can be measured with air ion spectrometers, but typical state-of-the-art condensation particle counters, which detect both neutral and charged clusters, only see particles larger than 2.5 nm in diameter. In this study we present a modified pulse-height condensation particle counter (PH-CPC) and confirm by laboratory verification that it is capable of detecting charged clusters with electrical mobility equivalent diameter down to ∼1 nm. We show how the detection efficiency and the pulse heights depend on the calibration particle size, polarity and composition. The effect of butanol supersaturation on the PH-CPC counting efficiency is also discussed. Furthermore, we developed an inversion method for the data to obtain true particle size distribution from the measurement signal.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.