Abstract
Introduction A robust genetic test for BRCA1 and BRCA2 genes is necessary for the diagnosis, prognosis, and treatment of patients with hereditary breast and ovarian cancer. We evaluated a commercial amplicon-based massively parallel sequencing (MPS) assay, BRCA MASTR Plus on the MiSeq platform, for germline BRCA genetic testing. Methods This study was performed on 31 DNA from cell lines and proficiency testing samples to establish the accuracy of the assay. A reference cell line DNA, NA12878 was used to determine the reproducibility of the assay. Discordant MPS result was resolved orthogonally by the current gold-standard Sanger sequencing method. Results The analytical accuracy, sensitivity, and specificity for variant detection were 93.55, 92.86, and 100.00%, respectively. Both sequencing depth and variant allele frequencies were highly reproducible by comparing the NA12878 DNA tested in three separate runs. The single discordant result, later confirmed by Sanger sequencing was due to the inability of the MASTR Reporter software to identify a 40-bp deletion in BRCA1 . Conclusion The BRCA MASTR Plus assay on the MiSeq platform is accurate and reproducible for germline BRCA genetic testing, making it suitable for use in a clinical diagnostic laboratory. However, Sanger sequencing may still serve as a confirmatory method to improve diagnostic capability of the MPS assay.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.