Abstract

The Sophysa Pressio (Sophysa Ltd., Orsay, France) is a new intracranial pressure monitoring system. This study aimed to evaluate its accuracy and compare it with the popular Codman intracranial pressure transducer (Codman/Johnson & Johnson, Raynham, MA) in vitro. A computerized rig was used to test the Pressio and Codman transducers simultaneously. Properties that were tested included drift over 7 days, the effect of temperature on drift, frequency response, the accuracy of measurement of static and pulsatile pressures, and connectivity of the system. Long-term (7 d) relative zero drift was less than 0.05 mmHg. The temperature drift was low (0.3 mmHg/207C). Absolute static accuracy was better than 0.5 mmHg over the range of 0 to 100 mmHg. Pulse waveform accuracy, relative to the Codman transducer, was better than 0.2 mmHg over the range of 1 to 20 mmHg. The frequency bandwidth of the Pressio transducer was 22 Hz. The Pressio monitor can transmit data directly to an external computer without the use of a pressure bridge amplifier. The new Pressio transducer proved to be accurate for measuring static and dynamic pressure during in vitro evaluation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.