Abstract
Venous cannulae undergo continuous improvements to achieve better and safer venous drainage. Several cannula tests have been reported, though cannula performance during inlet obstruction has never been a test criterion. In this study, five different cannulae for proximal venous drainage were tested in a mock circulation that enabled measurement of hydraulic conductance after inlet obstruction by vessel collapse. Values for hydraulic conductance ranged from 1.11 x 10(-2) L/min/mm Hg for a Thin-Flex Single Stage Venous Cannula with an open-end lighthouse tip to 1.55 x 10(-2) L/min/mm Hg for a DLP VAD Venous Cannula featuring a swirled tip profile, showing a difference that amounts to nearly 40% of the lowest conductance value. Excessive venous drainage results in potentially dangerous high-negative venous line pressures independent of cannula design. Cannulatip design featuring swirled and grooved tip structures increases drainage capacity and enhances cannula performance during inlet obstruction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.