Abstract
We report the results of a laboratory sensitivity and specificity evaluation of the Rapid Analyte Measurement Platform (RAMP®) Dengue Virus (DENV) antigen detection assay, which is designed to detect all serotypes of DENV in mosquito pools. The RAMP DENV assay was able to detect geographically distinct strains of all 4 DENV serotypes in virus-spiked mosquito pools that contained at least 4.3 log10 plaque forming units/ml, although discrete sensitivity limits varied slightly for each serotype. The RAMP DENV assay also detected DENV 1-4 in mosquito pools containing a single infected mosquito and 24 laboratory-reared uninfected mosquitoes. No false positives were detected in negative control mosquito pools or in samples containing high titers of nontarget arboviruses. We found that while the kit-supplied RAMP buffer reduced the infectious titer of DENV, it did not completely inactivate all serotypes. We recommend adding a detergent, Triton X-100, to the buffer to ensure complete inactivation of DENV if the assay is to be conducted at a lower biosafety level than required for DENV handling.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the American Mosquito Control Association
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.