Abstract

Pressure transducers are being used with increasing frequency in patient care. The recent availability of sterile transducer domes containing a diaphragm provides a means for reducing the risk of bacteremia when using these devices. The present studies measured the effect of diaphragm domes on the accuracy of pressure measurement. The sensitivity and frequency responses of 5 pressure transducers utilizing diaphragm and nondiaphragm domes were compared. The results indicate that the type of dome utilized and the method of dome application can significantly modify pressure transducer performance. The data are useful in providing optimal accuracy and reliability of pressure measurments when using a diaphragm dome, pressure transducer combination.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.