Abstract

We assessed the laboratory performance of the Chembio dual-path platform HIV-syphilis rapid immunodiagnostic test and electronic reader for detection of HIV and Treponema pallidum antibodies in 450 previously characterized serum specimens. For visual or electronic reader HIV antibody detection, the sensitivity was 100% and the specificity was 98.7%. For visual T. pallidum antibody detection, the test sensitivity was 94.7% and the specificity was 100.0%; with the electronic reader, the sensitivity was 94.7% and the specificity was 99.7%.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call