Abstract

Heat and moisture exchangers (HMEs) are used for airway humidification in mechanically ventilated patients and have been evaluated only under hospital conditions. U.S. Air Force aeromedical evacuation transports are performed under rugged conditions further complicated by the cold and dry environment in military aircrafts, and HMEs are used to provide airway humidification for patients. This study evaluated 10 commercial HMEs using a test system that simulated aeromedical evacuation conditions. Although the American National Standards Institute recommends inspired air to be at an absolute humidity value of > or = 30 mg/L for mechanically ventilated patients, the highest absolute humidity by any HME was approximately 20 mg/L. Although none of the HMEs were able to maintain a temperature high enough to achieve the humidity standard of the American National Standards Institute, the clinical significance of this standard may be less important than the relative humidity maintained in the respired air, especially on evacuation flights of short duration.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.