Abstract

Particle sensors offer significant advantages of compact size and low cost, and have recently drawn great attention for usage as portable monitors measuring particulate matter mass concentrations. However, most sensor systems have not been thoroughly evaluated with standardized calibration protocols, and their data quality is not well documented. In this work, three low-cost particle sensors based on light scattering (Shinyei PPD42NS, Samyoung DSM501A, and Sharp GP2Y1010AU0F) were evaluated by calibration methods adapted from the US EPA 2013 Air Sensor Workshop recommendations. With a SidePak (TSI Inc., St. Paul, MN, USA), a scanning mobility particle sizer (TSI Inc.), and an AirAssure™ PM2.5 Indoor Air Quality Monitor (TSI Inc.), which itself relies on a GP2Y1010AU0F sensor as reference instruments, six performance aspects were examined: linearity of response, precision of measurement, limit of detection, dependence on particle composition, dependence on particle size, and relative humidity and temperatu...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.