Abstract

Welding fume is a common exposure in occupational settings. Gravimetric analysis for total particulate matter is common; however, the cost of laboratory analyses limits the availability of quantitative exposure assessment for welding fume metal constituents in occupational settings. We investigated whether a field portable X-ray fluorescence spectrometer (FP-XRF) could provide accurate estimates of personal exposures to metals common in welding fume (chromium, copper, manganese, nickel, vanadium, and zinc). The FP-XRF requires less training and is easier to deploy in many settings than traditional wet laboratory analyses. Filters were analyzed both by FP-XRF and inductively coupled plasma mass spectrometry (ICP-MS). We estimated the FP-XRF limit of detection for each metal and developed a correction factor accounting for the non-uniform deposition pattern on filter samples collected with an Institute of Medicine (IOM) inhalable particulate matter sampler. Strong linear correlation was observed for all metals (0.72<r < 0.96). The median percent bias for chromium and nickel was less than 15%. The linear slope between the two methods for some metals (copper, manganese, and zinc) was greater than 1, indicating that the FP-XRF overestimated metal mass (median percent bias for vanadium was the largest at 94%), but the linearity of the response suggests that appropriate correction factors could be developed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.