Abstract

ABSTRACTThe Aerodyne Aerosol Chemical Speciation Monitor (ACSM) is well suited for measuring non-refractory particulate matter up to approximately 1.0 µm in aerodynamic diameter (NR-sub-PM1). However, for larger particles the detection efficiency is limited by losses in the sampling inlet system and through the standard aerodynamic focusing lens. In addition, larger particles have reduced collection efficiency due to particle bounce at the vaporizer. These factors have limited the NR-sub-PM1 ACSM from meeting PM2.5 (particulate matter with aerodynamic diameter smaller than 2.5 µm) monitoring standards. To overcome these limitations, we have redesigned the sampling inlet, the aerodynamic lens, and particle vaporizer. Both the new lens and vaporizer are tested in the lab using a quadruple aerosol mass spectrometer (QAMS) system equipped with light scattering module. Our results show that the capture vaporizer introduces additional thermal decomposition of both inorganic and organic compounds, requiring modi...