Abstract

The newly redesigned colorimetric VITEK-2 Compact system with updated Advanced Expert System (AES) (bioMerieux, Marcy l'Etoile, France) was evaluated for its accuracy and rapidity to identify clinical isolates and to detect several antimicrobial resistances. Overall, the VITEK-2 gave 95.8% of compatibility with the reference API strips (bioMerieux) in the identifications (IDs) of Gram-positive cocci (GPC), Gram-negative rods (GNR), and yeasts. The accuracy was finally estimated to 98.3% through additional confirmatory tests. Also, >90% of IDs of GPC and GNR were obtained within 7 h of incubations. The VITEK AES correctly detected 97.7% of antimicrobial resistances, including extended-spectrum β-lactamases, oxacillin and inducible clindamycin resistances in staphylococci, vancomycin resistance in enterococci, and penicillin and erythromycin resistances in Streptococcus pneumoniae. The most resistant isolates were identified within 12 h of incubations. In conclusion, the new colorimetric VITEK-2 Compact system with AES greatly improved its accuracy in species ID and detection of antimicrobial resistances, and it will be highly acceptable to clinical microbiology laboratory function.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.