Abstract
This paper presents a fully-integrated lab-on-a-chip (LOC) system for label-free detection and real-time counting of dielectrophoretically trapped multidrug resistant (MDR) K562 cells. The system integrates a parylene-based microfluidic DEP channel on top of a CMOS image sensor for the first time in the literature. The DEP channel can trap MDR K562 cells with 9 Vpp and 10 µl/min flow rate, and the CMOS image sensor can detect the trapped cells as small as 3 µm in diameter with a noise level of 28.3 e-rms.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.