Abstract
A facile, sensitive, and label-free assay for detection of interleukin-6 (IL-6) using liquid-gated field-effect transistor (FET) sensors based on horizontally aligned single-walled carbon nanotubes (SWCNT) is proposed. This approach relies on the drain current (Id) responses of the transistor upon interactions of IL-6 with its corresponding antibody (IL-6R) immobilized on SWCNT. The proposed immunosensor exhibits superior sensitivity (limit-of-detection=1.37pg/mL) in virtue of the reduced tube-to-tube contact resistance, good selectivity (no responses to bovine serum albumin and cysteine were observed and detection of target molecules in serum was achieved) as a result of the highly specific interaction between IL-6 and IL-6R, and excellent stability (no significant degradation in the electronic performance after storage under ambient conditions for up to 3months) in virtue of the strong adhesion of CNT to the quartz substrate and good horizontal alignment of these tubes. Therefore, the proposed immunosensor is a promising platform for early diagnosis of various diseases (including some cancers) that can be indicated by the circulating level of IL-6.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.