Abstract

Methods for the total content and individual determination of linear alkylbenzene sulfonates (LAS) in water samples based on the use of a lab-on-valve (LOV) module alone or coupled to CE equipment, respectively, have been developed. The total content of LAS has been determined by intrinsic absorption measurements (DA method) and after reaction with a methyl orange-cetylpyridine chloride mixture (MO method) with detection limits (LODs) of 21 ng/L and 15 microg/L, respectively, quantification limits (LOQs) of 70 ng/L, 50 microg/L, and development times of 100 and 124 s, respectively. The method for individual separation-quantification of LAS at very low concentration is based on automatic SPE preconcentration in the LOV module coupled on-line with the CE equipment. The LODs and LOQs thus obtained range between 1-21 and 4-70 ng/L, respectively, with linear dynamic ranges from the LOQ to 10 microg/L. Preconcentration factors of 10,000 and high efficiency to eliminate interferents by SPE enable application of the method to treated effluent, waste, surface and sea waters.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.