Abstract

We report here, for the first time, a smartphone-based lab-on-paper method for determination of uranium in ores. The method is based on the reaction with arsenazo-III on a filter-paper strip. The intensity of the green color was determined by a smartphone. The calibration curve is linear up to at least 60 μg mL−1 of uranium with LOD and LOQ values of 0.66 and 2.21 μg mL−1, respectively. The proposed method was applied for the analysis of spiked water samples and real Egyptian ore samples with recoveries ranging from 97.3 to 104.7%. The results are in agreement with the reported reference titrimetric method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.