Abstract

To evaluate the equivalency of the prototype of a new infrared ear thermometer (IRT 4000) in comparison to the temperature measurements of a pulmonary artery catheter (PAC) and a continual contact probe on the tympanic membrane. After approval by the local Ethics Committee, 26 (six female and 20 male) patients undergoing elective cardiac surgery were included in a single center open trial. During surgery, temperature measurements were recorded with the IRT 4000 in the left ear, and with a tympanic contact probe in the right ear, as well as with a PAC as reference. Measurements with the infrared ear thermometers and continual PAC values were recorded every six minutes. The average temperature measured with the IRT 4000 was 0.08 degrees C above the temperature of the PAC (95% confidence interval from -0.44-0.61 degrees C). Infrared ear thermometers 4000 temperature readings accurately reflect body core temperature and correlate well with the invasive PAC. Infrared ear thermometers may present a clinically useful alternative to the pulmonary artery thermometry for the measurement of core temperature in the perioperative setting or in the intensive care unit.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.