Abstract
Image-guided radiotherapy (IGRT) is an increasingly commonly adopted technique. As a result, however, total patient dose is increasing rapidly, especially when kV-cone beam computed tomography (CBCT) is applied. This study investigated the dosimetry of kV-CBCT using a Farmer ionization chamber with a (60)Co absorbed-dose calibration factor. The absorbed-dose measurements were performed using an I'mRT phantom (RW3, IBA) which is employed for dose verification of intensity-modulated radiotherapy (IMRT). The I'mRT phantom was used as a substitute for head and pelvis phantoms. The kV-CBCT absorbed dose was evaluated from a beam quality conversion factor of kV to (60)Co and the ionization ratio of the I'mRT phantom and water, calculated using the Monte Carlo method. The dose distribution in the I'mRT phantom was also measured using a radiophotoluminescent glass dosimeter (RGD). The absorbed doses for the pelvis phantom (full scan) ranged from 2.5-4 cGy for kV-CBCT and 4-8 cGy for MV-CBCT. TomoTherapy resulted in a lower dose of approximately 1.3 cGy due to fan-beam. For the head phantom (half scan), the doses ranged from 0.1-0.7 cGy for kV-CBCT and 3-5 cGy for MVCBCT. The results for RGD were similar to ion chamber measurements. It is necessary to decrease the absorbed dose of the organs at risk every time IGRT is applied.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.