Abstract
This date list mainly reports on samples processed from 1982 to 1991. Each sample is converted to methanol. Radiocarbon is measured in a Teflon™ vial containing a mixture of 40 g of methanol and 50 cc of xylene, 1% Butyl PBD and 0.1% PBBO using low-background scintillation counters Aloka LSC-800, LB-I and LB-III. The background of LB-III is ca. 8 cpm and counting efficiency is ca. 70%. The recent 14C standard used is 95% of NBS oxalic acid; 5568 yr is used as the half-life of 14C.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.