Abstract

In this study we evaluated the performance of the SHIGA TOXIN QUIK CHEK (Techlab®, Blacksburg, VA) and the ImmunoCard STAT! Enterohaemorrhagic E. coli (EHEC) (Meridian BioScience, Cincinnati, OH, USA) assays as methods for qualitatively detecting the presence of Shiga toxin in human fecal specimens. A multiplex PCR for the detection of stx1 and stx2 was used as the standard for comparison. The SHIGA TOXIN QUIK CHEK detected all known Shiga toxin subtypes with the exception of Stx2f, while the ImmunoCard STAT! EHEC was unable to identify four of the seven Stx2 subtypes, including Stx2b and Stx2d. When compared to multiplex PCR based on Shiga toxin gene presence alone both assays demonstrated 100% specificity, and gave sensitivity values of 50.0% and 41.2% respectively. Correlation between each assay and the multiplex PCR was calculated by the use of kappa, with both assays exhibiting a moderate level of agreement.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.