Abstract
Surface electromyography (sEMG) sensor is widely used to operate a myoelectric prosthesis and it has been further developed for various applications. In this paper, a multi-channel sEMG knit band sensor is newly suggested to be worn easily by amputee for myoelectric prosthesis. The band sensor was fabricated by knitting the silver-plated conductive yarn for electrodes with the moisture-wicking technical yarn for the comfort, ultimately, to record eight-channel sEMG signals simultaneously. Several experiments were conducted to verify the performance of knit band sensor. The normalized correlation between the developed knitted electrodes and the disposable (Ag/AgCl) electrodes is very high (0.95) at the same muscular area. The average signal-to-noise ratio value of the knitted electrodes (17.91 dB) was higher than that of the disposable electrodes (13.30 dB). For the performances of motion classification, overall classification accuracy of the knitted electrodes (93.2%) was higher than that of the disposable (87.9%). These experimental results have supported that the developed knit band sensor was as effective as the disposable electrodes.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.