Abstract

IntroductionInjury to the popliteal vessels during total knee replacement is rare but can lead to catastrophic outcomes.ReportAn 81 year old female presented with Rutherford IIb acute left limb ischaemia (ALI) 13 years after total knee replacement. The polyethylene insert in the knee implant had dislocated from the other components and had moved into the popliteal fossa, leading to popliteal artery compression. She underwent emergency multidisciplinary surgery including removal of the polyethylene component, thrombectomy, and popliteal artery stenting, but major amputation was required. The popliteal artery and the stent were removed and submitted to histological analysis. The stent was well expanded but focal malapposition was observed.ConclusionRegular follow up is mandatory in order to anticipate malfunction of the prosthesis and avoid long term complications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.