Abstract

The applicability of commercially available antisera for use in kinetic immunoturbidimetry has been studied using the protein albumin as a simple model. The sensitivity of the kinetic immunoturbidimetric approach has been found to be comparable to published data for nephelometric systems, being able to detect concentrations of protein as low as 1 mg/l. The technique has shown to be as precise as most dye binding techniques for the measurement of albumin, producing a within batch C.V. of 1.0% and a between batch C.V. of 1.5%. The technique was found to be comparable to another immunological technique (RID). The fast assay time (30–60 sec), cost and good precision makes this the method of choice for routine albumin measurement. The extreme sensitivity of the technique of kinetic immunoturbidimetry makes the technique applicable to the estimation of a wide range of proteins in blood, urine and CSF.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.