Abstract

AbstractAn optical chemical sensor (optode) is proposed for the kinetic determination of nitrite ion. The optode was fabricated by the immobilization of methyl violet on a triacetylcellulose polymeric membrane. Methyl violet is covalently bonded to a transparent triacetylcellulose film. By immersion of the sensor into an acidic nitrite solution, the absorbance of the sensor at 596 nm decreases with time that is due to the reaction of nitrite with the immobilized methyl violet. A fixed time method of 15 min was used to monitor the reaction. The linear range for the determination of nitrite was 0.20–8.00 µg mL−1 and the limit of detection was 0.08 µg mL−1. The optodes were one‐shot, they had durability more than 2 months and were easily prepared. The optode was successfully applied to the determination of nitrite ion in spring water and sewage samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.