Abstract

Silver ion can perturb the Belousov-Zhabotinskii (B-Z) oscillating chemical reaction. Therefore, the B-Z oscillating system was applied in the determination of silver ion by using a platinum wire as an indicator electrode in the potentiometric method. The amplitude of the potentiometric oscillation increased linearly in proportion to [Ag+] in the range of 9.42 x 10(-6) M to 2.54 x 10(-4) M, with a correlation coefficient of 0.999 under the optimum conditions. The obtained LOD (2sigma) was 8.85 x 10(-6) M and the relative standard deviation (RSD) for five measurements of 1 x 10(-4) M silver ion was 5%. The influence of some potentially interference was also investigated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.