Abstract
Purpose: Compare percent depth dose (PDD) and off-center ratio (OCR) measured by the CC13 ionization chamber and the RAZOR silicon diode in small photon beams. Method and Materials: Some dosimetric characteristics, such as PDD, OCR, penumbra and radiation field size, were considered in this study for 2x2, 3x3, and 4x4 cm2 field sizes. We used the CC13 ionization chamber and the RAZOR silicon diode to measure dose distribution with depth along the axis and off-center of the beam. From the results obtained, the team investigated the differences in radiation parameters measured by the two types of probes above. Results: There are significant differences in the radiation parameters investigated for the CC13 ionization chamber and the RAZOR silicon diode, especially the width of penumbra. For PDD curves, the difference is less than 5% from dmax to 30 cm, however the difference becomes greater in the build-up region, which reaches to 33% at the water phatom surface. The width of penumbra measured by CC13 is always larger than that of RAZOR, the ratio of the penumbra width between two detectors is 1.8 and 1.3 for energies of 6 MV and 15 MV, respectively. Conclusion: The RAZOR silicon diode has better dose response than the CC13 ionization chamber for measuring the PDD and the OCR in small photon beams.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Clinical Medicine- Hue Central Hospital
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.