Abstract

A comparison of the dosimetry for accelerator photon beams was carried out between the Dutch Metrology Institute (VSL) and the Bureau International des Poids et Mesures (BIPM) from 23 September to 20 October 2014. The comparison was based on the determination of absorbed dose to water for three radiation qualities of the medical accelerator facilities of the National Physical Laboratory (United Kingdom). After establishing Draft B, the VSL discovered an error in the calculation of the correction factor for excess-heat linked to the VSL glass vessel used in the measurements at the NPL. The comparison results for the revised standard, reported as ratios of the VSL and the BIPM evaluations (and with the combined standard uncertainties given in parentheses), are 0.9959 (54) at 6 MV, 0.9958 (64) at 10 MV and 0.9991 (75) at 25 MV. This result is part of the on-going BIPM.RI(I)-K6 series of comparisons.Main textTo reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.