Abstract

Photoplethysmography (PPG) is an optical technique for detection of blood volume changes in the microvascular bed of a biological tissue. Many aspects of the PPG signal formation are still unclear. In particular, it is not known how the shape of a registered PPG signal depends on the geometry of tissue illumination. The aim of this study is to model the PPG waveform using the Monte Carlo (MC) method. For this, we developed a three-layer optical model of the skin in a reflectance geometry and verified it experimentally for different wavelengths (660, 810, and 940 nm) and source-detector distances (from 4 to 10 mm). The MC simulation results showed that the PPG waveform depends on the source-detector distance. The most pronounced diastolic wave is observed at the distance of 6 mm for the wavelength of 810 nm. The results obtained can be used for the development of reflectance PPG sensors.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call