Abstract
Recent studies showed that J waves were associated with higher incidence of ventricular tachyarrhythmia (VT/VF) in patients with idiopathic ventricular fibrillation (VF) and myocardial infarction. We sought to assess the association between J waves and VT/VF in patients with nonischemic cardiomyopathy (NICM). We retrospectively enrolled 109 patients (79 men; mean age, 60 ± 15 years) with NICM who underwent implantable cardioverter defibrillator (ICD) implantation. The primary endpoint of this study was the occurrence of appropriate device therapy due to sustained VT/VF. The J wave was electrocardiographically defined as an elevation of the terminal portion of the QRS complex of >0.1 mV in at least 2 contiguous inferior or lateral leads. Among the 109 patients, 37 (34%) experienced an episode of appropriate device therapy during a median follow-up period of 25.9 (IQR 11.5-54.3) months. Kaplan-Meier curves showed that the presence of J waves on the 12-lead ECG obtained before device implantation was associated with an increased occurrence of appropriate device therapy (P < 0.001). Multivariate Cox proportional regression analysis revealed that the presence of J waves (HR 2.95; 95% CI 1.31-6.64; P = 0.009) was an independent predictor for the occurrence of appropriate device therapy. In the subgroup analysis of the patients with dilated or hypertrophic cardiomyopathy, J wave tended to increase the occurrence of appropriate device therapy (P = 0.056 and P = 0.092, respectively). The presence of J waves was an independent predictor for the occurrence of appropriate device therapy in patients with NICM who underwent ICD implantation.