Abstract

Light-pipe radiation thermometers (LPRTs) are the sensor system of choice in RTP tools. They can be calibrated against blackbodies with an uncertainty (k=1) less than 0.3 °C. In an RTP tool, however, account must be made for wafer emissivity and wafer-chamber interreflections, or else temperature measurement uncertainties will be orders of magnitude higher. We have used two complementary approaches for accomplishing this: 1) in situ calibration using high-accuracy wire/thin-film thermocouples calibrated on the International Temperature Scale of 1990 (ITS-90) and 2) developing optical models to estimate the effective emissivity of the wafer eeff when used in the radiation environment of the RTP tool. The temperature measurement uncertainty of LPRTs using either technique is 2.1 °C or less.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.