Abstract
Modern advancements in radiation therapy require paralleled advancements in the dosimetric tools used to verify dose distributions. Optical computed tomography (CT) imaged radiochromic gel dosimeters provide comprehensive, tissue equivalent, 3D dosimetric information with high spatial resolution and low imaging times. Traditional CT image reconstruction methods (filtered backprojection) do not account for light refraction within the optical CT system reducing the image quality. Iterative reconstruction methods make use of a system matrix that describes this light refraction thus, improving the reconstructed image quality. However, use of iterative reconstruction methods is not widespread, largely due to the impractical storage size of the required system matrix. Furthermore, current iterative reconstruction methods do not address the issue of image degradation due to a single detector element collecting light from multiple raypaths. For optical CT radiochromic gel dosimetry to be used effectively as a radiation therapy treatment plan verification tool, the system must be both practical and accurate. Thus, this work has two main objectives: (i) reduce the size of system matrices by means of polar coordinate discretization in lieu of the traditional Cartesian coordinate discretization, and (ii) reduce image degradation due to multiple raypaths by a novel approach to populating the system matrix that accounts for multiple raypaths.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.