Abstract
Criterion-related validity of the thermodilution cardiac output technique for cardiac output measurement has to have a high correlation (r = .91 to .98) with the direct Fick method, the gold standard of cardiac output measurement. Issues that can affect validity of the measurements include the position of the pulmonary artery catheter, the rate of injection of the indicator solution, the volume and temperature of the injectate, the timing of the injection of indicator solution during the respiratory cycle, the position of the subject, and the presence of concomitant infusions. Variation in measurement can be limited by considering the delivery system for the indicator solution, by recording time-temperature cardiac output curves, and by considering normal biologic variations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.