This work reports on the development of a lab demonstrated resonant mass sensor towards mid-size production. The issues associated with scaling-up production of the microfabricated chip are discussed with particular focus on yield and device reproducibility, as well as the constraints imposed on the design and manufacturing of the device when packaging and integration must be taken into account. Issues of modal alignment and ambient operational pressure are discussed. Fabricated devices show a 4.81 Hz pg−1 mass sensitivity with a temperature sensitivity of typically 10 Hz °C−1.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.