Abstract
The use of inductively coupled plasma atomic emission spectrometry (ICP-AES) for the determination of 235U: 238U isotope ratios in U–Zr metal alloys (90% m/m U, 10% m/m Zr) is described. Conventional pneumatic nebulization and laser ablation sample introduction techniques were utilized. The results of the determination agreed, within experimental uncertainty, with isotope ratios determined by thermal ionization mass spectrometry (TIMS), e.g., 235U: 238U for conventional nebulization = 2.091 ± 0.006, for laser ablation = 2.092 ± 0.015 and for TIMS = 2.0940 ± 0.0004. The precision, i.e., the relative standard deviation (RSD), of the sequential determination of the intensities of the 235U and 238U components of the U emission line and the resultant isotope ratios was improved significantly by the use of an intrinsic internal standard (Zr), i.e., RSD = 1.6% to RSD = 0.17%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.