Abstract

This study identified the reason for the poor quantification of lormetazepam-TMS (1) using negative ion chemical ionization with lormetazepam-1,1,1-2H3-TMS (2) as an internal standard. Mass spectra of lormetazepam and its deuterium labelled compounds determined at various ion source temperatures (100-250 degrees C) gave almost the same behaviour for 1 and lormetazepam-3',4',5',6'-2H4-TMS (3) but a different one for 2, suggesting that the poor quantification was due to an isotope effect. This was confirmed by the findings that the ratios of ion currents of the base peaks of 1 and 3 were independent of the ion source temperature but those of 1 and 2 varied markedly with it. This phenomenon was also observed in the mass fragmentography of the molecular ion, although to a lesser degree than that of the above fragment. In both positive ion chemical ionization and electron impact ionization modes, no isotope effect arose because there was no corresponding fragment to cause the isotope effect.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.