Abstract
A direct and sensitive method for the determination of blood cyanide by isotope dilution was developed. The blood is placed in a headspace vial, and K13C15N is added as internal standard. Addition of phosphoric acid liberates the cyanide as HCN. The detection is accomplished by mass spectrometry after a fine mass calibration tuning. The detection limit obtained is 0.3 micromol/L. The within- and inter-run coefficients of variation are 4.4% (for a concentration of 2.5 micromol/L) and 3.9% (for a concentration of 4.7 micromol/L), respectively. The observed recovery is 98%. A round-robin exercise was carried out to compare the performance of this method with others currently in use in other clinical laboratories.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.